exposure maintenance formula calculator

Increasing the kVp increases IR exposure and the density produced on a film image, and decreasing the kVp decreases IR exposure and the density produced on a film image (Figure 10-4). Cost Per Unit. Math Application 10-2 The radiographer must decide how much of a change in mAs is needed to correct for the density error. Select Distance Units Select Exposure Rate Units mR/hr Feet Feet mR/hr Calculated Final Dose-Rate Enter Starting Dose-Rate Enter Ending Distance Enter Starting Distance Inverse Square Law Calculations source-to-object distance (SOD) Assuming that the anatomic part is adequately penetrated, changing the kVp does not affect the digital image the same as a film-screen image. WebLong Exposure Calculator app screen shot showing a baseline shutter speed of 1/25th of a second and the changes to the shutter speed depending on what ND filter I choose at the top. Want to know how much you should set aside to cover the cost of warranties. It equals a net exposure of 60%. WebWhat is the formula for the Magnification Factor? You maintain density by increasing mAs as distance also increases. When the radiograph is deemed unacceptable, this means the optical densities lie outside the films sensitometric curves straight-line portion, and may need to be repeated. The 15% rule states that changing the kVp by 15% has the same effect as doubling the mAs, or reducing the mAs by 50%; for example, increasing the kVp from 82 to 94 (15%) produces the same exposure to the IR as increasing the mAs from 10 to 20. Because kVp affects the amount of radiation reaching the IR, its effect on the digital image is similar to the effect of mAs. Maintenance margin is the amount of equity, expressed as a percentage, that must be maintained in a margin account. Increasing the kVp increases the penetrating power of the radiation and increases the exposure to the image receptor. Equation mAs1 = (D1) squared mAs2 (D2) squared 15% rule There are 3 good reasons to use the 15% rule Decrease in scatter Decrease in patient dose Decrease time/wear & tear on the tube When using the 15% rule, you are determining beam quantity: "the total number of x-ray photons in a beam." You will, therefore, have to pay a higher premium if you are a frequent driver, commensurate with your higher loss exposure. To increase the mAs to 20, you could use: As demonstrated in Math Application 10-1, mAs can be doubled by doubling the milliamperage or doubling the exposure time. If your exposure/histogram already looks good but you would like to modify ISO, aperture, or shutter speed, just dial in your current settings and check its new value depending on changes applied to the other two. If a radiograph must be repeated because of another error, such as positioning, the radiographer may also use the opportunity to make an adjustment in density to produce a radiograph of optimal quality. When using a film-screen IR, radiographers need to assess the level of density produced on the processed image and determine whether the density is sufficient to visualize the anatomic area of interest. Critical Concept 10-3 Recognize how to correct exposure factors for a density error. Calculate changes in mAs when adding or removing a grid. This is why the exposure dependence on (s) is also linear as it is with mA. An. Radiographic Exposure Technique Product Enquiry. exposure route. As mentioned earlier, the effects of changes in the kVp are not uniform throughout the range of kVp. Adequate penetration of the anatomic area of interest is equally important when using digital IRs, and therefore kVp selection is important in producing a quality image. inverse square law i 1 (D2)2 ----- = ------- i 2 (D1)2 Get more background on exposure and camera settings from Fred Parker . The mAs does not have a direct effect on image brightness when using digital IRs. Critical Concept 10-7 The film characteristic, speed, and chemical processing determine the amount of optical density produced on the image for a given mAs. **Disclaimer: CREFCOA's yield maintenance calculator is used to help estimate a potential prepayment penalty based on a yield maintenance prepayment formula. Similarly, too little radiation reaching the IR (within reason) produces a digital image with the appropriate level of brightness, but the increased noise decreases image quality. Exposure Rating = 12 weeks * 25% probability = 4 weeks risk exposure Interested in understanding more Lean-Agile Principles? You can also check what the exact shutter speed should be if you add a Neutral Density (ND) filter. Focal Spot Size The voltage potential used to generate the x-rays (KeV). All you have to do is choose the settings for the aperture f-stop, shutter speed, and ISO sensitivity. Depending on how far above ground the RF source is located, you might want to consider ground reflections and then click "Calculate". Knowledge of how these factors affect the image individually and in combination assists the radiographer to produce a radiographic image with the amount of information desired for diagnostic interpretation. The size of the electron stream depends on the size of the filament. B, A greater decrease in mAs, one-fourth the original mAs, is needed. Adjusting Milliamperage and Exposure Time to Maintain mAs The surgical procedure would take about 3 hrs. Radiographs that have sufficient but not optimal density usually are not repeated. State exposure technique modifications for the following considerations: body habitus, pediatric patients, projections and positions, soft tissue, casts and splints, and pathologic conditions. Show all work. In addition, whenever a 15% change is made in the kVp to maintain the exposure to the IR, the radiographer must adjust the mAs by a factor of 2. WebThe formula is based off of the assumption that hospitalized patients have greater energy expenditure and determines fluid requirements based on weight alone (a proxy for energy expenditure in a non-linear relationship). In general, for repeat radiographs necessitated by density errors, the mAs is adjusted by a factor of 2; therefore a minimum change involves doubling or halving the mAs. 1) Daily volume formula: (100 mL for each of the first 10 kg) + (50 mL for each kg between 11 and 20) + (20 mL for each additional kg past 20 kg) = 1,000 mL + 500 mL + 120 mL = 1,620 mL. A 15% decrease in kVp has the same effect as decreasing the mAs by half. However, regardless of the type of IR, the patient should be exposed to the least amount of radiation necessary to produce a diagnostic-quality radiograph. As mentioned previously, it may take more than doubling the mAs to correct for a density error. When low or high kVps are used, the amount of change in the kVp required to maintain the exposure to the IR may be greater or less than 15%. Risk is calculated by multiplying the impact or "value" of a loss with its frequency or probability of occurring. Want to know how much you should set aside to cover the cost of warranties. Making a visible change in radiographic density requires that the minimum amount of change in mAs be approximately 30% (depending on equipment, this may vary between 25% and 35%). Pathologic Conditions Kilovoltage The formula for women is: BMR (kcal / day)= 10 weight (kg) + 6.25 height (cm) 5 age (y) 161 (kcal / day) For men, it is: Three days later a surgeon, wishing to operate, hesitates to do so fearing radiation exposure. ionizing radiation. WebReview List the three methods of reducing your exposure/dose: Intensity decreases _____ with the square of the distance from the source due only to the change in _____. Use decimal notation for percentage, i.e., 2% is .02. 80kVp1.15=92kVp I wrote this tool to plan long exposures outside my camera's typical range. Show all work. direct square law formula maintains. In order to calculate the net exposure, you subtract 20% from 80%. A greater increase is needed for high kVp (90 and above) than for low kVp (below 70). WebUse the Exposure Maintenance Formula. WebUsed to determine how to maintain density when SID is changed. Although the level of brightness has been adjusted, there may be increased quantum noise visible within the image. State exposure technique modifications for the following considerations: body habitus, pediatric patients, projections and positions, soft tissue, casts and splints, and pathologic conditions. Shutter Speed. This means that the actual production time has Altering the penetrating power of the x-ray beam affects its absorption and transmission through the anatomic tissue being radiographed. When a small focal spot is used, the heat created during the x-ray exposure is concentrated in a smaller area and could cause tube damage. When the radiographer selects a particular focal spot size, he or she is actually selecting a filament size that is energized during x-ray production. The quantity of radiation reaching the patient affects the amount of remnant radiation reaching the IR. The smaller this stream, the greater the heat generated in a small area; therefore it is desirable to have a larger actual focal spot area. What are the correction factors used for 3 phase 6 pulse & 3 phase 12 pulse? *Interest calculations based on 30/360 day calendar year. The probability of Compton scattering is related to the energy of the photon. In other words, if I double my mAs value, the radiation exposure value will double. A patients age, condition, and the presence of a pathologic condition also affect the amount of mAs required for the procedure. Assuming that the anatomic part is adequately penetrated, changing the kVp does not affect the digital image the same as a film-screen image. However, the kVp has a greater effect on the image when using film-screen IRs. If not calculating based on ideal body weight, use clinical judgment for dosing. Assuming that the anatomic part is adequately penetrated, changing the kVp does not affect the digital image in the same way as a film-screen image. As x-ray photon energy increases, the probability of that photon penetrating a given tissue without interaction increases. WebThe scale allows one to make ratio comparisons on intensity ratings and to determine the magnitude of direct. Focal spot size is an important consideration for the radiographer because the focal spot size affects recorded detail. Fill in the following form to determine whether a workers heat stress is above recommended limits. As a result, images with lower contrast are produced. . WebWHAT IS THE EXPOSURE MAINTENANCE FORMULA? WebWhile originally derived in pediatric patients, this calculator is applicable to any age. The area of interest must be adequately penetrated before the mAs can be adjusted to produce a quality radiographic image. Cn Tn)8 E is the equivalent exposure for the working shift. WebExample of a IV fluid calculation. Milliamperage and Exposure Time Ingestion exposure can occur via consumption of contaminated food, water and other liquids. Formula: mAs1 (D1)2 ------- = ------- mAs2 (D2)2 WITH CHANGES IN SID, WHAT FORMULA IS USED TO CALCULATE THE NEW EXPOSURE RATE? Lets continue with the same example. The selection of kVp alters its absorption and transmission through the anatomic part regardless of the type of IR used and therefore must be selected wisely. What % increase is the intensity of the beam? When a small focal spot is used, the heat created during the x-ray exposure is concentrated in a smaller area and could cause tube damage. If not calculating based on ideal body weight, use clinical judgment for dosing. Formula: mAs1 (D1)2 ------- = ------- mAs2 (D2)2 WITH CHANGES IN SID, WHAT FORMULA IS USED TO CALCULATE THE NEW EXPOSURE RATE? As mentioned earlier, the effects of changes in the kVp are not uniform throughout the range of kVp. Make the Physics Connection 10-5 Although the level of brightness has been adjusted, there may be increased quantum noise visible within the image. Body Habitus **Disclaimer: CREFCOA's yield maintenance calculator is used to help estimate a potential prepayment penalty based on a yield maintenance prepayment formula. To maintain exposure to the IR, when increasing the kVp by 15% (kVp 1.15), divide the original mAs by 2. WebThe formula is based off of the assumption that hospitalized patients have greater energy expenditure and determines fluid requirements based on weight alone (a proxy for energy expenditure in a non-linear relationship). This calculator should not be relied upon for final decision making. Repeated exposures made just under the limit over a long period can still jeopardize the life of the x-ray tube. Show all work. WebClear. You can do that with the Mifflin-St Jeor formula that requires weight, height, age, and sex. If the radiograph necessitates an adjustment greater than a factor of 2, the radiographer should multiply or divide the mAs by 4 (Figure 10-3). MF = Image Size/Object Size: What is the formula for Heat Units (HU) for single phase? The primary exposure technique factors the radiographer selects on the control panel are milliamperage, time of exposure, and kilovoltage peak (kVp). Calculate the magnification factor and determine image and object size. When a film image is too light (insufficient density), a greater increase in mAs may be needed to correct the density, or the mAs may need to be decreased to correct a film image that has excessive density. It is important for the radiographer to determine the amount of mAs needed to produce a diagnostic image. A, Larger focal spot. WebA patient has been given 5 mCi Ga-67 citrate for imaging infection. The physical dimensions of the focal spot on the anode target in x-ray tubes used in standard radiographic applications usually range from 0.5 to 1.2 mm. WebBelow is the formula used by OSHA to determine the cumulative exposure for an eight-hour shift: E = (Ca Ta+Cb Tb+. (From Fauber TL: Radiographic imaging and exposure, ed 3, St Louis, 2009, Mosby.) Chapter 9 emphasizes that a good-quality radiographic image accurately represents the anatomic area of interest. A high kVp results in less absorption and more transmission in the anatomic tissues, which results in less variation in the x-ray intensities exiting the patient (remnant), producing a low-contrast (long-scale) image. ionizing radiation. WebThis is How to Calculate Warranty Liability Costs. Special Considerations Excessive Radiation Exposure and Digital Imaging For film-screen IRs, the mAs controls the density produced in the image. Man hours is the total hour worked over a specific period of time. contrast medium Logarythmic exposure calculator slide rules are commonly used to quickly and efficiently calculate radiographic exposure times based on current sealed sources activity and thickness of material. Key Terms To simplify the calculation, we always assume that 2 10 is equal to 1000 instead of 1024 since it is negligibly small. WebPrevention Heat Stress Calculator. Focal spot size is determined by the filament size. WebExposure Calculations Properly exposing a radiograph is often a trial and error process, as there are many variables that affect the final radiograph. 5 5. Get more background on exposure and camera settings from Fred Parker . Equation mAs1 = (D1) squared mAs2 (D2) squared 15% rule There are 3 good reasons to use the 15% rule Decrease in scatter Decrease in patient dose Decrease time/wear & tear on the tube When using the 15% rule, you are determining beam quantity: "the total number of x-ray photons in a beam." A greater change in the kVp is needed when operating at a high kVp (greater than 90) compared with operating at a low kVp (less than 70) (Figure 10-5). Consider using ideal body weight in obese patients. To calculate your maintenance calories: Count your BMR - basal metabolic rate. Chapter 10 Also, the original mAs and the original distance are both in the numerator direct vs. inverse. We need to be using this formula when the question asks for a mAs value. 200mA50ms(0.05s)=10mAs kVp and mAs Conversely, when the quantity of x-rays is decreased, the exposure to the IR decreases. 3 phase 6 pulse 1.35 (35%) & 3 phase 12 pulse 1.41 (41%). Film-Screen Speed To best visualize the anatomic area of interest, the mAs selected must produce a sufficient amount of radiation reaching the IR, regardless of type. When adequate penetration is achieved, further increasing the kVp results in more radiation reaching the IR. Explain how kVp affects radiation production and image receptor exposure. Recognize the factors that affect recorded detail and distortion. A 15% increase in kVp has the same effect as doubling the mAs. H-117 Introductory Health Physics Slide 31 Using the inverse square law, calculate the dose rate at 4 feet away from a point source if the dose rate is originally 1000 R/hr at 2 feet. body habitus How to calculate TEEP. For film-screen IRs, the mAs controls the density produced in the image. Because kVp affects the amount of radiation reaching the IR, its effect on the digital image is similar to the effect of mAs. Additionally, for a given mAs, IRs respond differently. WebExposure Calculator. Useful wikipedia articles: Exposure Value. Focal spot size influences the amount of unsharpness recorded in the image. Maintenance margin is the amount of equity, expressed as a percentage, that must be maintained in a margin account. When low or high kVps are used, the amount of change in the kVp required to maintain the exposure to the IR may be greater or less than 15%. Unlike mAs, the kVp affects the amount of radiation exposure to the IR and radiographic contrast. Estimates are given in mrem, the U.S. unit for effective dose. Compare the effect of changes in milliamperage and exposure time on film-screen and digital images. This is why the exposure dependence on (s) is also linear as it is with mA. 5 5. During computer processing, image brightness is maintained when the mAs is too low or too high. Radiographs that have sufficient but not optimal density usually are not repeated. How to calculate TEEP. WebThis calculator only estimates your child maintenance payments - it cannot tell you exactly what youll get. However, the kVp has a greater effect on the image when using film-screen IRs. In these interactions, more Compton scattering than x-ray absorption occurs (photoelectric effect) and more scatter exits the patient. Small focal spot sizes are usually 0.5 or 0.6 mm, and large focal spot sizes are usually 1 or 1.2 mm in size. WebCalculate? It is important for the radiographer to determine the amount of mAs needed to produce a diagnostic image. ASTM Standards: Consensus Standards For Systems, Services, Materials And Product. . You maintain density by increasing mAs as distance also increases. Recognize patient factors that may affect image receptor exposure. Exposure techniques using higher kVp with lower mAs exposure techniques are recommended in digital imaging because contrast is primarily controlled during computer processing. How To Calculate Man Hours. Depending on the type of control panel, milliamperage and exposure time may be selected separately or combined as one factor, milliamperage/second (mAs). How To Calculate Man Hours. kVp and the Radiographic Image WebThis points towards the requirement of proper scheduling of preventive maintenance activities, so that the work load is even for all days. Changing the kVp affects its absorption and transmission as it interacts with anatomic tissue; however, using a higher kVp reduces the total number of interactions and increases the amount of x-rays transmitted. Please read OSHAs Heat Hazard Recognition page for more information about WBGT, workload, acclimatization status, and clothing. Weight. 6- stops extends my shutter speed to 2.5 seconds. Use decimal notation for percentage, i.e., 2% is .02. This chapter focuses on radiographic exposure techniques and the use of accessory devices, and their effect on the radiation reaching the image receptor (IR) and the image produced. WebFor 10-stops ND filter, if the base shutter speed (without filter) is seconds, with the use of 10-stops ND filter, the exposure time becomes x 2 10 = x 1000 = 250 seconds. Fill in the following form to determine whether a workers heat stress is above recommended limits. WebExample of a IV fluid calculation. As focal spot size increases, unsharpness increases and recorded detail decreases; as focal spot size decreases, unsharpness decreases and recorded detail increases. Many secondary factors affect the radiation reaching the IR and image quality. The complexities of Quantitative Risk Analysis When decreasing the kVp by 15% (kVp 0.85), multiply the mAs by 2. The level of radiographic contrast desired, and therefore the kVp selected, depends on the type and composition of the anatomic tissue, the structures that must be visualized, and to some extent the diagnosticians preference. Logarythmic exposure calculator slide rules are commonly used to quickly and efficiently calculate radiographic exposure times based on current sealed sources activity and thickness of material. T is the duration in hours of the exposure at the concentration C. At higher kVp, a greater proportion of Compton scattering occurs compared with x-ray absorption (photoelectric effect), which decreases radiographic contrast. As a result, images with lower contrast are produced (Figure 10-6). This relationship between radiation exposure intensity and density is discussed in more detail in Chapter 9. FIGURE 10-4 kVp and Radiation Exposure.Increasing the kVp increases the penetrating power of the radiation and increases the exposure to the image receptor. All you have to do is choose the settings for the aperture f-stop, shutter speed, and ISO sensitivity. For example, when the mAs is increased, density is increased; when the mAs is decreased, density is decreased (Figure 10-2). Explain the relationship between milliamperage and exposure time with radiation production and image receptor exposure. WebEVS = EV100 + log2(S / 100) where S = desired ISO. magnification factor (MF) 3 phase 6 pulse 1.35 (35%) & 3 phase 12 pulse 1.41 (41%). kVp does not produce the same effect across the entire range of kVp used in radiography. Typically, exposure occurs by one of three exposure routesinhalation, ingestion, or dermal. Why Use. The effective focal spot is the origin of the x-ray beam and is the area as seen from the patients perspective. Make the Physics Connection 10-1 mAs must increase when distance increases in order to maintain IR exposure. As focal spot size changes, so does the amount of unsharpness. How to use this calculator. These factors make achieving a desired level of radiographic contrast more complex than achieving a desired level of radiographic density, especially for film-screen imaging. This typically brings the optical densities back within the straight-line portion of the films sensitometric curve to best visualize the anatomic area of interest. the mAs (technique)needed to maintain optimal exposure at varying SIDs. The distance between the source of the radiation and the IR, source-to-image receptor distance (SID) Typically, exposure occurs by one of three exposure routesinhalation, ingestion, or dermal. Cost Per Unit. Equation mAs1 = (D1) squared mAs2 (D2) squared 15% rule There are 3 good reasons to use the 15% rule Decrease in scatter Decrease in patient dose Decrease time/wear & tear on the tube When using the 15% rule, you are determining beam quantity: "the total number of x-ray photons in a beam." The kVp affects the exposure to the IR because it alters the amount and penetrating ability of the x-ray beam. 80kVp0.85=68kVp Remember that a 15% change in kVp does not produce the same effect across the entire range of kVp used in radiography. Tags: Essentials of Radiographic Physics and Imaging Changing the kVp affects its absorption and transmission as it interacts with anatomic tissue; however, using a higher kVp reduces the total number of interactions and increases the amount of x-rays transmitted. The kVp affects the contrast in a digital image; however, image brightness and contrast are primarily controlled during computer processing. Source-to-Image Receptor Distance Compare the effect of changes in milliamperage and exposure time on film-screen and digital images. ionizing radiation. When using a film-screen IR, radiographers need to assess the level of density produced on the processed image and determine whether the density is sufficient to visualize the anatomic area of interest. Kilovoltage is not a factor typically manipulated to vary the amount of IR exposure in film-screen imaging because the kVp also affects contrast. HU = mA x Sec x kV. A few months later, the average downtime per day was only 20 minutes. In order to calculate the net exposure, you subtract 20% from 80%. Man hours is the total hour worked over a specific period of time. Modern radiographic x-ray generators are equipped with safety circuits that prevent an exposure from being made if that exposure exceeds the tube loading capacity for the focal spot size selected. The actual focal spot is the area being bombarded by the filament electrons. The smaller this area of origin, the sharper the image. Critical Concept 10-2 When the kVp is increased at the control panel, a larger potential difference occurs in the x-ray tube, giving more electrons the kinetic energy to produce x-rays and increasing the kinetic energy overall. Maintaining or adjusting exposure to the IR can be accomplished with kVp by using the 15% rule. 100mA0.2s=20mAs Z= gives us the straining of the outer fiber. In general, the smallest focal spot size available should be used for every exposure. Kilovoltage, Scatter Radiation, and Radiographic Contrast. mAs must increase when distance increases in order to maintain IR exposure. In general, the smallest focal spot size available should be used for every exposure. The characteristics evaluated for image quality are density or brightness, contrast, recorded detail or spatial resolution, distortion, and noise. mAs must increase when distance increases in order to maintain IR exposure. WebUse the Exposure Maintenance Formula. WebFor 10-stops ND filter, if the base shutter speed (without filter) is seconds, with the use of 10-stops ND filter, the exposure time becomes x 2 10 = x 1000 = 250 seconds. Describe the use of grids and beam restriction, and their effect on image receptor exposure and image quality. mAs and Digital Image Brightness If the mAs selected is too high (high exposure to the digital IR), the brightness can also be adjusted, but the patient has received more radiation than necessary. Decreasing the kVp decreases the proportion of Compton scattering and increases radiographic contrast. WebCalculate? As the time of the exposure is increased by a factor of 2, the number of x-ray photons coming out of the x-ray tube will also be increased by a factor of 2. Radiographic images generally are not repeated to make only a slight visible change. F-Number (Aperture) Sunny f/16 rule. Calculate changes in milliamperage and exposure time to change or maintain exposure to the image receptor. ASTM Standards: Consensus Standards For Systems, Services, Materials And Product. Log In or Register to continue WebThe scale allows one to make ratio comparisons on intensity ratings and to determine the magnitude of direct. Example of a Margin Call An investor is looking to purchase a security for $100 with an initial margin of 50% (meaning the investor is using $50 of his money to purchase the security and borrowing the remaining $50 from a Calculate changes in milliamperage and exposure time to change or maintain exposure to the image receptor. (From Fauber TL: Radiographic imaging and exposure, ed 3, St Louis, 2009, Mosby.) WebBelow is the formula used by OSHA to determine the cumulative exposure for an eight-hour shift: E = (Ca Ta+Cb Tb+. Identify the exposure factors that can affect patient radiation exposure. The 200,000 figure in the formula represents the number of hours 100 employees working 40 hours per week, 50 weeks per year would work, and provides the standard base for calculating incidence rate for an entire year. Critical Concept 10-1 In 2015, the New York state legislature enacted a major revision to the maintenance law. X 300mAs 47" 60" This problem has been solved! Critical Concept 10-5 It equals a net exposure of 60%. Recognize the factors that affect recorded detail and distortion. Critical Concept 10-9 Calculate changes in mAs for changes in source-to-image receptor distance. Additionally, when a kVp that is too high is selected, the image brightness and contrast are adjusted, but patient exposure may be increased. Estimate the radiation dose to the surgeon, and compare it with common radiation exposure sources. The film characteristic, speed, and chemical processing determine the amount of optical density produced on the image for a given mAs. When to Use. Please read OSHAs Heat Hazard Recognition page for more information about WBGT, workload, acclimatization status, and clothing. FIGURE 10-5 kVp Range and Radiographic Density.Produced at 50 kVp (A) and produced at 90 kVp with the mAs adjusted to maintain radiographic density (C). After detecting that one of the biggest issues with the factory was too much downtime, the maintenance manager took steps to improve that matter. AKA the exposure maintenance formula: as SID increases, intensity decreases, causing a decrease in film exposure and density. Lets continue with the same example. WebThe time of the exposure is a relatively easy concept to understand. The voltage potential used to generate the x-rays (KeV). 85 mAs 225 mAs 55" 3. Increasing the kVp by 15% increases the exposure to the IR, unless the mAs is decreased. D= represents the depth of Calculating Magnification

Uil Spirit State Championship 2022 Pictures, Urban League Columbus Ohio Housing List, Articles E

exposure maintenance formula calculator